用于医疗设备的电池组

发布时间:2013-02-16 16:37:53

用于医疗设备的电池组,可循环使用2000次

 

No.

Item

Specification

 

Remarks

1

Nominal Capacity

9Ah

23±3℃,discharge to 8V cut off, Min.9Ah

2

Nominal Voltage

12V

Standard Output Nominal Voltage

3

Standard Charging Current

2A

On 23℃ ,2 CC(constant current)changed to 14.6V,then CV(constant voltage)charge till charge

current decline to ≤0.05C

4

Max. Charging Current

5A

On 23℃,5A CC(constant current )discharger cut-off .

6

Standard Discharging Current for battery

2A

On 23℃2A CC(constant current )discharger to 8V cut-off .

7

Charge Ending Voltage

14.6V

Best for 2A CC/CV 14.6V, 0.05C cut off.

8

Discharge Ending Voltage

8V

2.3V for each cell,For the battery pack, it will be

over than 8V.

 9

Max. Discharging Current

15A

On 23℃, The battery working discharge keeps 10A.

10

Instant Max.

Discharging Current

30A

On 23℃, battery instant Max discharging current

less than 30A.

11

Temperature while charging

0℃45℃

Battery charging process , battery and ambient

temperature does not exceed 45℃

12

Temperature while

discharging

-20℃60℃

The specified temperature range, the battery work normally within the framework of capacity loss in the error

14

Battery Weight

About 1Kg

Including wires and packaging material

15

Packaging

 

Shrink by PVC.

 